The main purpose of a 5m semi anechoic chamber is to perform full compliance measurements of radiated emission and immunity at 5m or 3m test distance. The combination of the pan-type 2mm galvanized panel system, parallel closing RF shielded doors and polystyrene hybrid absorbers on the walls, floor and ceiling creates a high performance controlled electromagnetic environment that complies with international immunity and emission test standards.
The main purpose of a 5m semi anechoic chamber is to perform full compliance measurements of radiated emission and immunity at 5m or 3m test distance. The combination of the pan-type 2mm galvanized panel system, parallel closing RF shielded doors and polystyrene hybrid absorbers on the walls, floor and ceiling creates a high performance controlled electromagnetic environment that complies with international immunity and emission test standards.
Features of a 5m semi anechoic chamber:
Full compliance (5m) emission
test site for frequency range of 30MHz to 18GHz as per CISPR 16-1-4 (2010) +annex A1 (2012) (free space testing)
Full compliance (3m) emission
test site for frequency range of 30MHz to 18GHz as per CISPR 16-1-4 (2010) +annex A1 (2012) (free space testing)
Full compliance immunity test
site for frequency range of 80MHz to 6GHz as per IEC/EN 61000-4-3
Shielding effectiveness
according to EN 50147-1 (>80dB at 10kHz)
Acceptance standards for a 5m semi anechoic
chamber:
CISPR 16-1-4
IEC/EN 61000-4-3
EN 50147-1
ANSI C63.4
FCC part15
CISPR 25(special customization)
MIL-STD-461(special
customization)
Product standards for a 5m
semi anechoic chamber:
CISPR 32
CISPR 35
CISPR 11
CISPR 14
CISPR 15
CISPR 12
GB/T18387
Typical performance of a 5m semi anechoic
chamber:
FSNSA: +/-4dB at 5m test distance; +/-3.5dB at 3m test distance
sVSWR: ≤ 5.5dB at 3m test distance
FU: 75% of 16 points ≤ 6dB at 3m test distance & 1.5 x 1.5m test window (0.8m
above floor level)
Typical specifications for a 5m semi anechoic chamber:
External dimensions |
12.7m*7.6m*6.6m(L*W*H) |
Dimensions |
12.1m*7.0m*6.0m(L*W*H) |
Control room dimensions |
5.0m*4.0m*3.0m(L*W*H) |
Amplifier room dimensions |
3.0m*3.0m*3.0m(L*W*H) |
Construction |
2mm galvanized shielding panels with a free standing structural steel support |
Absorbers |
Hybrid solution of ferrite tiles and polystyrene hybrid absorber on walls, floor and ceiling |
Ventilation |
300mm*300mm / 18GHz |
Shielded door |
RF shielded swing door 1.2m*2.1m(W*H) |
Ground interface board |
400mm*400mm/18GHz, including RF connectors |
EUT power line |
Filtered 220V 32A 50Hz/60Hz; 380V 63A 50Hz/60Hz |
Lighting |
Interference-free LED automatic lifting and lighting system |
Frequency range |
30MHz to 18GHz for radiated emission and 80MHz to 6GHz for radiated immunity |
Quiet zone size |
Diameter 2m, height 3m |
Positioning system |
Optical fiber control turntable, turntable diameter 3m; Optical fiber control antenna tower, 1-4m lifting. |
In addition to the typical specifications of 5m semi anechoic chamber, RFT can also customize the anechoic chamber configuration for customers according to the size of quiet zone and the type of test standard.
5m semi anechoic chamber option
After special design of anechoic chamber, CISPR25 test table and matching grounding system are installed, CISPR25 emission and immunity test are carried out, and the original performance of anechoic chamber would not be affected.